Simulates respiration at 8 rates (3 to 150 BrPM), 11 impedance variations (0.1
to 10 Ohms) and 11 basal impedances (100 to 4000 Ohms). Apnea in
INS/EXP.
Selected Information:
Calibrated control of 8 respiration rates, 11 impedance variations and 11
basal impedances.
Impedance variation is almost independent of excitation frequency from dc to
100 kHz.
Simulate apnea in either inspiration or expiration. Manually generate any
desired respiration rate.
Impedance output connects directly to monitor or can be used with an ECG
Simulator.
Battery power or external power source.
Certificate of Conformance traceable to NIST is
available.
The M98C Respiration Simulator is used to
quantitatively test monitors that detect changes in thoracic impedance. It
dynamically simulates 11 impedance variations over a wide range of basal
impedances and respiration rates. Apnea can be simulated either in expiration or
inspiration independent of the respiratory rate setting.
Since the
Respiration Simulator is a floating, purely resistive source, simple dc
measurement instruments can be used to verify the accuracy of simulated
impedance changes as well as basal impedances.
When testing monitors that
detect both ECG and respiration, the M98C can be connected in series with either
the Fogg M310 or M311 ECG Simulator. Their low output resistance has a minimal
effect on simulated basal impedance and they feature selection of four ECG
shapes and a wide range of heart rates and amplitudes.